XNR-400C Melt Flow Rate Tester (Melt Index Meter)
1、 Product Name: Melt Flow Rate Tester (Melt Index Meter)
2、 Product model: XNR-400C3、 Product Usage: (Volume Method, Quality Method)
4、 Product Introduction:
The XNR-400C melt flow rate tester (melt index meter) meets the standard requirements of GB/T3682-2000. The main parameters of the melt index meter also meet the requirements of ISO1133-97, ASTM1238, JIS and other standards. It can be used for testing polyethylene, polypropylene, polystyrene ABS、 The melt mass flow rate (MFR) or melt volume flow rate (MVR) of various plastic materials such as polyamide, fiber resin, acrylic ester, polyoxymethylene, fluoroplastics, and polycarbonate are widely used in plastic production, plastic products, petrochemical industry, as well as related universities, research institutions, and inspection departments for measurement.
5、 Product features:
Ø High precision Chinese and English color controller, specially customized film buttons, have unparalleled advantages in control accuracy and service life
The integrated control of manual and automatic cutting is precise to 0.1 seconds, and the cutting time and number of cutting times can be set arbitrarily.
Ø Built in micro printer, capable of printing test data
Ø Dual temperature control system, using dual inlet heating coils, dual inlet PT100 platinum resistance sensors, independent temperature control for upper and lower sections, improving testing accuracy and temperature stability.
Add displacement sensors to test MVR (volume method) with an accuracy of 0.01mm.
6、 Testing methods: Quality method, Volume method (MFR, MVR)
7、 Display mode: Chinese and English color LCD display
8、 Main technical parameters:
1. Temperature range: 0 ℃ -400 ℃
2. Temperature fluctuation: ± 0.2 ℃
3. Temperature uniformity: ± 1 ℃
4. Temperature display resolution: 0.1 ℃
5. Time display resolution: 0.1S
6. Barrel diameter: Φ 2.095 ± 0.005mm
7. Discharge port length: 8.000 ± 0.025mm
8. Diameter of loading cylinder: Φ 9.550 ± 0.025mm
9. Weight accuracy: ± 0.5%
10. Output method: Micro automatic printing output
11. Cutting method: Hand automatic cutting integrated
12. Test load: 8 levels in total, 8 sets of weights
13. Power supply voltage: AC220V ± 10%, 50HZ